MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPT; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number B35200 |
Device Problems
Break (1069); Impedance Problem (2950)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
Injury
|
Event Description
|
Information was received from a manufacturer¿s representative (rep) reported the original implanting physician didn¿t use a burr hole device, but rather used a cranial plating system to secure the lead and it ¿basically wore through the lead causing an impedance issue.¿ the patient was currently having the lead and extension removed due to the impedance issues and was going to place a new lead and extension the following day (the battery was going to remain implanted and in use and was therefore plugged until the following day).
|
|
Manufacturer Narrative
|
Section d information references the main component of the system.Other relevant device(s) are: product id b3400040m lot# serial# (b)(6).Implanted: explanted: product type extension product id b3300542 lot# serial# (b)(6).Implanted: explanted: product type lead section d information references the main component of the system.Other relevant device(s) are: product id: b3400040m, serial/lot #: (b)(6), ubd: , udi#: (b)(4).; product id: b3300542, serial/lot #: (b)(6) , ubd: 14-jun-2024, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from the manufacturer¿s representative (rep), which was confirmed with the healthcare provider (hcp), reported following lead and extension replacement the impedance issue resolved.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
B5: all previously applied coding is only relevant for the lead.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|