ETHICON ENDO-SURGERY, LLC. OPT BLADELESS 5X100 STABILITY; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Catalog Number 2B5LT
Device Problems Crack (1135); Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 10/3/2023.D4: batch # unk.Additional information was requested and the following was obtained: "please provide more details about ¿became stuck on the end of the trocar sleeve, causing damage¿ they were unsure how damage could have occurred such as through residual heat from the l hook or if the l hook was accidentally activated during removal.The l hook itself appeared to be damaged / bent prior to the issue with the trocar per the surgical team.Did the trocar sleeve break off in patient? it did not appear that anything broke off." investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that only the sleeve from the 2b5lt device was received with the tip of the sleeve damaged melted.The event reported was confirmed and it is related to improper use of the device.In order to prevent this kind of damage please avoid the contact with laser, electrosurgical or ultrasonic devices.As part of ethicon¿s quality process all devices are manufactured, inspected, and released to approved specifications.The manufacturing records couldn't be reviewed as the batch number is unknown.
 
Event Description
It was reported that during a laparoscopic cholecystectomy, when removing the laparoscopic monopolar l hook it became stuck on the end of the trocar sleeve, causing damage.It was later noted that the l hook was bent prior to getting stuck on the trocar.No delays.Procedure was completed.No patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPT BLADELESS 5X100 STABILITY
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key17861124
MDR Text Key324802578
Report Number3005075853-2023-07121
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10705036013709
UDI-Public10705036013709
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2B5LT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2023
Initial Date Manufacturer Received 09/06/2023
Initial Date FDA Received10/03/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-