ETHICON ENDO-SURGERY, LLC. OPT BLADELESS 5X100 STABILITY; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
|
Back to Search Results |
|
Catalog Number 2B5LT |
Device Problems
Crack (1135); Use of Device Problem (1670)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/06/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Date sent: 10/3/2023.D4: batch # unk.Additional information was requested and the following was obtained: "please provide more details about ¿became stuck on the end of the trocar sleeve, causing damage¿ they were unsure how damage could have occurred such as through residual heat from the l hook or if the l hook was accidentally activated during removal.The l hook itself appeared to be damaged / bent prior to the issue with the trocar per the surgical team.Did the trocar sleeve break off in patient? it did not appear that anything broke off." investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that only the sleeve from the 2b5lt device was received with the tip of the sleeve damaged melted.The event reported was confirmed and it is related to improper use of the device.In order to prevent this kind of damage please avoid the contact with laser, electrosurgical or ultrasonic devices.As part of ethicon¿s quality process all devices are manufactured, inspected, and released to approved specifications.The manufacturing records couldn't be reviewed as the batch number is unknown.
|
|
Event Description
|
It was reported that during a laparoscopic cholecystectomy, when removing the laparoscopic monopolar l hook it became stuck on the end of the trocar sleeve, causing damage.It was later noted that the l hook was bent prior to getting stuck on the trocar.No delays.Procedure was completed.No patient harm.
|
|
Search Alerts/Recalls
|
|
|