MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE X3; PATIENT MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE X3; PATIENT MONITOR Back to Search Results
Model Number 867030
Device Problem Device Fell (4014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/12/2023
Event Type  malfunction  
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Event Description
The customer reported that the x3 fell and had damage.The device was not in use on a patient at the time of event, there was no patient involvement.
 
Manufacturer Narrative
A philips field service engineer (fse) went onsite to evaluate the device in question.The fse confirmed the unit fell and it was damaged outside of use.The unit was not connected to a patient monitor at the time of the event.Based on the information available and the testing conducted, the cause of the reported problem was confirmed to be user handling.After the handle and handle pin were replaced, the device was returned to functional use with no further issues identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE X3
Type of Device
PATIENT MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key17861792
MDR Text Key324813567
Report Number9610816-2023-00497
Device Sequence Number1
Product Code MHX
UDI-Device Identifier00884838082588
UDI-Public00884838082588
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K171801
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number867030
Device Catalogue Number867030
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/12/2023
Initial Date FDA Received10/03/2023
Supplement Dates Manufacturer Received10/03/2023
Supplement Dates FDA Received10/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-