Catalog Number 383942
Device Problem Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2023
Event Type  malfunction  
Event Description
It was reported that while using the bd pegasus¿ safety closed iv catheter system the needle would not disengage.The following was translated from chinese to english: the head nurse reported that after implanting an indwelling needle in the patient, she found that the needle core could not be withdrawn.
 
Manufacturer Narrative
E.1.Initial reporter phone #: (b)(6).H.6.Investigation summary: in response to the event reported by your facility a device history review was conducted for lot number 3135911.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.A sample could not be obtained for evaluation and testing; in lieu of the affected device, functional testing was performed on retention samples for this lot, the results of these show that the tested units performed within product specifications.Unfortunately without the ability to investigate the affected unit our quality engineers were unable to determine the root cause for this complaint.Bd will continue to monitor this issue and encourages you to submit your sample for review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer (Section G)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17865276
MDR Text Key324879594
Report Number3014704491-2023-00618
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number383942
Device Lot Number3135911
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/15/2023
Initial Date FDA Received10/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-