• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CUSTOM-PAK SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CUSTOM-PAK SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 16644-08
Device Problem Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that cannulas with white corrosion and clogged orifice occurred during calibration of cataract surgery with intraocular lens (iol) implant.There was no patient harm.
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.No physical sample has been returned for evaluation, therefore, the condition of the product could not be verified.When this type of issue occurs, a sample will need to be returned so that a proper investigation can be conducted.If possible, the customer should provide the company with a photo of the reported issue in order to help with the investigation of this complaint.A review of the reported pak's bill of materials (bom) shows there are two cannulas inside the pack: 8065420120- cannula,ant chamber,27g and oa4036j- cannula,hydrodis(chang),27gx3.The root cause of the customer's complaint could not be established as a sample has not been received and the condition of the product could not be verified.Without analysis of the sample, it is not possible to isolate the root cause.The root cause for this complaint is not known, therefore, specific action with regards to this complaint cannot be taken.Based on current tracking, there are no adverse trends for this reported complaint.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM-PAK SURGICAL PROCEDURE PACK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17867267
MDR Text Key324866341
Report Number1644019-2023-01341
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00000000000000
UDI-Public00000000000000
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number16644-08
Device Lot NumberJZ5182256
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/07/2023
Initial Date FDA Received10/04/2023
Supplement Dates Manufacturer Received11/28/2023
Supplement Dates FDA Received12/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ALCON MICROSURGICAL INSTRUMENTS CANNULA OPHTHALMIC
-
-