Device Problem Computer Software Problem (1112)
Patient Problem Insufficient Information (4580)
Event Date 09/11/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation was completed by conducting a thorough evaluation of the product and the reported information.The customer reported that two fields were accidentally delivered to the previously treated left hip instead of the prescribed right hip.Based on the analysis of the software logs two fields to the right hip were fully treated and were within the tolerances defined for the fields parameters.The treatment was recorded and mosaic was working as designed and intended.However, the customer reports that the fields were clinically delivered to the left hip in error.The mistreatment is due to use error.Due to use error radiation was delivered to an incorrect treatment site and elekta physics have assessed the use error as serious.
 
Event Description
The customer reported that two fields were accidentally delivered to the previously treated left hip instead of the prescribed right hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIQ
Type of Device
ACCELERATOR, LINEAR, MEDICAL
Manufacturer (Section D)
ELEKTA SOLUTIONS AB
kungstensgatan 18
box 7593
stockholm, SE-10 3 93
SW  SE-103 93
Manufacturer (Section G)
ELEKTA SOLUTIONS AB
400 perimeter center terrace
suite 50
atlanta GA 30346
Manufacturer Contact
cornerstone
london road
crawley, west sussex RH10 -9BL
MDR Report Key17867345
MDR Text Key324866213
Report Number3015232217-2023-00062
Device Sequence Number1
Product Code IYE
UDI-Device Identifier00858164002367
UDI-Public(01)00858164002367(10)2.83.040
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183034
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/13/2023
Initial Date FDA Received10/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/30/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-