|
Model Number 2404-02 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/02/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluated by mfr.: the device was returned for analysis.Visual inspection of the device revealed numerous kinks throughout the body of the wire.The wire was separated at the seam weld 38cm proximal of the tip.The separated/detached ends appeared to be bent, indicating that the device was severely kinked prior to separation.The sensor was visible and in the correct location within the sensor housing.The wire was gently shaken to see if the sensor would move and be more visible within the sensor housing.The sensor did not move which verified the sensor was not detached from the fiber optic cable.As it was not reported that the device was detached/separated and the shaft had numerous kinks, it was most likely the shaft/body was separated sometime after the procedure.
|
|
Event Description
|
Reportable based on device analysis completed on (b)(6) 2023.It was reported that no pressure signal occurred.The comet ii pressure guidewire was used for a diastolic hyperemia-free ratio (dfr)/fractional flow reserve (ffr) study.During the procedure, the comet ii pressure guidewire failed to connect, and a low signal strength led light (color yellow) was shown on the ffr link.The physician attempted a second time, but the connection still failed.The physician's decided to continue with the study the next day.The procedure was completed successfully the next day using a new comet ii pressure guidewire.There were no patient complications reported and the patient condition was stable.However, device analysis revealed a separated/detached wire.
|
|
Search Alerts/Recalls
|
|
|