ETHICON ENDO-SURGERY, LLC. LIGACLIP*MCA MED APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number MCM20
Device Problems Mechanical Problem (1384); Failure to Form Staple (2579)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 10/4/2023 d4: batch # x96g4w additional information was requested and the following was obtained: "please clarify how ¿the clips are not clamping or closing¿? the clips are not clamping and closing.Did device not fire clips (jammed)? no.Did device fire malformed clips? no.Did device fire scissored clips? no.Did clip not hold in tissue or vessel once placed? no." investigation summary the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the mcm20 device was received with no damage to the external components.Upon cycling, the instrument was noted to be empty and locked out.The instrument is designed to lockout after all the clips have been fired; therefore a potential cause for the customer reported experience is the firing of all of the clips, as a result, the instrument could no longer be fire due to the activation of the lockout mechanism.As the device was returned empty for evaluation we are unable to investigate further the issue of ¿multiple feed" and ¿malformed clips".The device was disassembled to verify the condition of the internal components and no anomalies were noted.  the event described could not be confirmed as the device was returned empty.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Event Description
It was reported that during an unknown procedure, clips are not clamping or closing and sometimes it is double clipping.
 
Manufacturer Narrative
(b)(4).Date sent: 11/9/2023.D4: batch # unk.Additional information was requested and the following was obtained: " were there any patient consequences? if yes, please describe.No".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key17868721
MDR Text Key324893309
Report Number3005075853-2023-07193
Device Sequence Number1
Product Code GDO
UDI-Device Identifier10705036002475
UDI-Public10705036002475
Combination Product (y/n)N
Reporter Country CodeTC
PMA/PMN Number
K820837
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMCM20
Device Lot Number253C00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/10/2023
Initial Date FDA Received10/04/2023
Supplement Dates Manufacturer Received10/11/2023
Supplement Dates FDA Received11/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/18/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-