Siemens has completed the investigation.The customer provided instrument log files, automatic quality control (aqc) files and sensor files.A review of this data indicates that the potassium sensor was performing as intended at time of analysis on the escalated samples.The instrument did not generate any error messages related to potassium sensor performance.It should be noted that pre-analytics such as differences in matrices (serum vs whole blood), collection practices, sample inhomogeneity, and time differences between comparative measurements may play a part of the observed difference between the results.The measurement cartridge was replaced, and the customer is operational.Although a definitive root cause could not be determined, there is no indication that the instrument is not operating as intended.
|