• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX500H11
Device Problems Degraded (1153); Moisture or Humidity Problem (2986)
Patient Problems Headache (1880); Unspecified Respiratory Problem (4464)
Event Date 09/12/2023
Event Type  malfunction  
Event Description
The manufacturer was contacted in reference to the voluntary field safety notice/recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging visualization of particles in the air path.The patient alleges congestion and headache.There was no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.The patient was contacted and confirmed seeing particles in the tank and moisture in the hose.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
H3 other text: device not returned to manufacturer.
 
Manufacturer Narrative
The manufacturer previously reported information in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging congestion and headache.The patient did not state whether medical intervention was required."other" was inadvertently selected under adverse event on the initial report.A pms clinical expert review made a determination based on information available at the time of this review, that the allegations do not constitute a serious injury, therefore the adverse event entry of "other" has been removed.The device has not yet been returned to the manufacturer for evaluation.The customer responded via e-mail on 09/20/2023 and stated the device is available to be returned for evaluation.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
6501 living place
pittsburgh PA 15208
Manufacturer (Section G)
RESPIRONICS, INC.
6501 living place
pittsburgh PA 15208
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key17869420
MDR Text Key325365771
Report Number2518422-2023-25062
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX500H11
Device Catalogue NumberDSX500H11
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/12/2023
Initial Date FDA Received10/04/2023
Supplement Dates Manufacturer Received10/02/2023
Supplement Dates FDA Received10/11/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/12/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1974-2021
Patient Sequence Number1
Patient Outcome(s) Other;
-
-