There was an allegation of questionable elecsys troponin t hs assay results for 13 patient samples on a cobas e 801 analytical unit.For sample 1, the initial result was 3.66 pg/ml.The repeat result was 19.9 pg/ml.For sample 2, the initial result was 4.75 pg/ml.The sample was repeated on an e411 analyzer and the repeat results were 1251 pg/ml and 1193 pg/ml.For sample 3, the initial result was 5.72 pg/ml.The repeat result was 12.6 pg/ml.For sample 4, the initial result was 4.33 pg/ml.The repeat result was 13.0 pg/ml.For sample 5, the initial result was 4.90 pg/ml.The repeat result was 10.1 pg/ml.For sample 6, the initial result was 4.29 pg/ml.The repeat result was 114 pg/ml.For sample 7, the initial result was 3.25 pg/ml.The repeat result was 316 pg/ml.For sample 8, the initial result was 5.34 pg/ml.The repeat result was 112 pg/ml.For sample 9, the initial result was 3.52 pg/ml.The repeat result was 14.0 pg/ml.For sample 10, the initial result was 5.26 pg/ml.The repeat result was 28.5 pg/ml.For sample 11, the initial result was 4.11 pg/ml.The repeat result was 7.24 pg/ml.For sample 12, the initial result was 4.01 pg/ml.The repeat result was 10.3 pg/ml.For sample 13, the initial result was 6.99 pg/ml.The repeat result was 26.8 pg/ml.The repeat results were deemed correct.No questionable results were reported outside of the laboratory.
|
It was found that the customer did not perform qc when the reagent pack was first installed.Qc was only performed when there were (b)(4) tests remaining on (b)(6) 2023.Per product labeling, "controls for the various concentration ranges should be run individually at least once every 24 hours when the test is in use, once per cobas e pack, and following each calibration." based on the available data, a general reagent issue could be excluded.The investigation did not identify a product problem.The root cause of the event could not be determined.
|