Catalog Number 306547 |
Device Problem
Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/21/2023 |
Event Type
malfunction
|
Event Description
|
It was reported while using the bd posiflush¿ normal saline syringe the plunger was difficult to move.The following was received from the initial reporter: customer seem to have certain lot number of prefilled flush syringes that the plunger is somewhat ¿stuck¿ to the barrel.It takes a fair amount of pressure to dislodge the plunger causing fluid to squirt from the syringe.
|
|
Manufacturer Narrative
|
B.3.Date of event: unknown.The date received by manufacturer has been used for this field.H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Manufacturer Narrative
|
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 11-oct-2023.H6: investigation summary: it was reported the plunger is stuck to the barrel.To aid in the investigation, one empty sample with no packaging flow wrap or tip cap was received for evaluation by our quality team.A visual inspection was performed, and no defects or imperfections were observed.The sample was then tested for sustaining force, which is the force applied to the plunger rod while moving downwards when expelling the saline solution, and the result was within specification.A device history record review was completed for provided material number 306547, lot 3207548.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.There were no related quality notifications.All processes and final inspections complied with specification requirements.Based on the investigation and with the returned sample analysis the symptom reported by the customer could not be confirmed and a probable root cause could not be determined.
|
|
Event Description
|
It was reported while using the bd posiflush¿ normal saline syringe the plunger was difficult to move.The following was received from the initial reporter: customer seem to have certain lot number of prefilled flush syringes that the plunger is somewhat ¿stuck¿ to the barrel.It takes a fair amount of pressure to dislodge the plunger causing fluid to squirt from the syringe.
|
|