• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. URETERO-RENO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P6
Device Problems Loose or Intermittent Connection (1371); Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus, the uretero-reno fiberscope had abnormal image and the instrument channel port was loose.The issue was found during preparation before use inspection.There were no reports of patient or user harm associated with this event.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegations of ¿abnormal image and the instrument channel port was loose¿ were not confirmed.The device evaluation found image guide bundle had significant breakages, watertightness was lost due to the bending section cover had pinhole damage and the adhesive was chipped.Due to wear of angle wire, bending angle in up direction did not meet the standard value.Due to damage, the angulation lever did not move smoothly.And the connecting tube had a scratch.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 5 years since the subject device was manufactured.Based on the results of the investigation, a specific root cause of the suggested event could not be established.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17873868
MDR Text Key324964995
Report Number9610595-2023-14622
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170340802
UDI-Public04953170340802
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K172298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/07/2023
Initial Date FDA Received10/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/04/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-