RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX500T11C
Device Problem Degraded (1153)
Patient Problems Excessive Tear Production (2235); Sneezing (2251); Unspecified Respiratory Problem (4464)
Event Date 08/16/2021
Event Type  malfunction  
Event Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The patient alleges nasal/throat irritation, runny eyes and sneezing.There is no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.The device was returned and evaluated by the manufacturer.The internal part of the device was inspected visually.The device's downloaded logs were reviewed by the manufacturer and 0 error was found.The manufacturer concludes that there was no evidence of visible foam degradation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key17876272
MDR Text Key324982778
Report Number2518422-2023-25332
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX500T11C
Device Catalogue NumberDSX500T11C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2022
Initial Date Manufacturer Received 08/16/2021
Initial Date FDA Received10/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1974-2021
Patient Sequence Number1
-
-