Model Number X SERIES |
Device Problem
No Display/Image (1183)
|
Patient Problem
Injection Site Reaction (4562)
|
Event Type
malfunction
|
Event Description
|
Complainant alleged that during a biomed testing, the device's screen turned completely white and was illegible.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical united kingdom for evaluation.The customer's report was observed and attributed to user mishandling.During disassembly of the device to determine root cause, the technician observed damage consistent with mis-handling.The observed damage was not consistent with normal wear and tear.The lcd display was replaced to resolve the report.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
|
Search Alerts/Recalls
|