Model Number DBEC-125
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Chest Pain (1776); Perforation of Vessels (2135)
Event Date 09/06/2023
Event Type  Injury  
Manufacturer Narrative
H6 investigation conclusion code 22: the diamondback 360® coronary orbital atherectomy system instructions for use manual states that vessel perforation and chest pain are potential adverse events that may occur and/or require intervention with use of the system.The device history record for this oad lot number has been reviewed.No issues or discrepancieswere noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.Csi id: (b)(4).
 
Event Description
A diamondback 360 coronary orbital atherectomy device (oad) was used to treat a right coronary artery lesion.The physician attempted to advance the viperwire advance guide wire with glide assist mode, but it was unable to advance.The oad was advanced through with a low-speed treatment.After three treatments, the patient experienced chest pain and angiographic imaging showed a perforation.A stent was placed to stop the bleeding.Balloon angioplasty was performed.Angiographic imaging showed no additional complications and the procedure was complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM
Type of Device
CORONARY ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
aaron stevens
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key17878027
MDR Text Key325007361
Report Number3004742232-2023-00250
Device Sequence Number1
Product Code MCX
UDI-Device Identifier10850000491530
UDI-Public(01)10850000491530(17)240630(10)437249-1
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P130005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 10/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberDBEC-125
Device Catalogue Number7-10060-05
Device Lot Number437249-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/10/2023
Initial Date FDA Received10/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age65 YR
Patient SexMale
-
-