Catalog Number MCM20 |
Device Problem
Failure to Form Staple (2579)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/19/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4) date sent: 10/5/2023 d4: batch # unk attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
It was reported that during a liver resection procedure, had to fire three times for placement of one clip.The direction of the clip was right or left , so there was misalignment.Replaced the instrument and continue the operation.No patient consequences.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 11/9/2023; d4: batch # a9cv2t.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the mcm20 device was returned with a clip in jaws and with no apparent damage.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the device was cycled and it fed and formed the remaining 10 clips as intended.Although no conclusion could be reach on the cause of the reported event, the instructions for use do contain the following caution: ensure that a clip is in the jaws.Position the jaws so that the clip completely surrounds the vessel to be ligated.Fully squeeze the handles together until they stop.As the handles are squeezed, the clip closes and occludes the vessel or tubular structure.Fully release the handles to return the instrument to its original open position.The next clip is automatically advanced.Inspect the jaws after each clip application to ensure the presence of a new clip before applying the next clip.Repeat steps as necessary to continue ligation of tissue.For optimal performance, periodically clean the instrument during the procedure by submerging the jaw and distal end of the applier shaft in saline.The event described could not be confirmed as the device performed without any difficulties noted.As part of ethicon¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 12/19/2023.Additional information was requested and the following was obtained: ""did the surgeon fire one device three times (clips fired right or left) and then changed to a second instrument, which worked well or did the surgeon used three different devices, where device one and two fired right or left clips and the third one worked well." answer: "the surgeon fire one device three times (clips fired right or left) and then changed to a second instrument, which worked well"" attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: "please clarify how the statement "the direction of the clip was right or left , so there was misalignment".Did the device feed slowly, sideways, or does not feed/advance into jaws? did the device multiple or double feed? did the device fire clips that are not formed completely or forms a pear/tear drop shape? did the device fire clips that the legs do not close parallel or cross? this may also include ribbon shaped or x-shaped.".
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 1/9/2024.Additional information was requested and the following was obtained: 1.Did the device feed slowly, sideways, or does not feed/advance into jaws? no.2.Did the device multiple or double feed? no.3.Did the device fire clips that are not formed completely or forms a pear/tear drop shape? no.4.Did the device fire clips that the legs do not close parallel or cross? this may also include ribbon shaped or x-shaped.Yes.
|
|