EDWARDS LIFESCIENCES COMMANDER DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9750CM26A
Device Problems Burst Container or Vessel (1074); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2023
Event Type  malfunction  
Event Description
During a tavr procedure using a 26mm sapien 3 ultra valve via transfemoral approach, the balloon rupture during valve deployment due to heave sinotubular junction (stj) calcium.A new delivery catheter was used and re-inflated valve keeping the balloon more on the inflow portion of valve at nominal inflation.No patient complications reported.The device will not be returned.
 
Manufacturer Narrative
The investigation is ongoing.H3 other text : device was discarded.
 
Manufacturer Narrative
The commander delivery system was not returned to edwards lifesciences for evaluation.Without the device returned for evaluation, visual inspection, functional testing and dimensional analysis were unable to be completed.A device history records review did not reveal any issues that could have contributed to the reported events.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of the failure mode is not required at this time.The complaint for balloon burst was unable to be confirmed as no device or relevant imagery provided.However, no manufacturing non-conformance was identified during the evaluation.An existing edwards technical summary has been documented for root cause analysis on balloon bursts in a calcified landing zone.The technical summary provides a rationale as to why it is unlikely that a product defect or manufacturing non-conformance contributed to this type of event, including factors on why deployment of balloons on thv delivery systems are subject to increased risk of burst in a calcified landing zone.The complaint description states, 'the balloon rupture during valve deployment due to heavy stj calcium'.The presence of calcification can create a challenging anatomy for balloon inflation.While the balloons are sufficiently designed and tested to ensure the burst pressure is at or above the rated burst pressure, calcified nodules can compromise the structure of the balloon wall via following mechanisms such as puncture, local overstretching, open cell impingement, or stress concentration.During manufacturing, the device undergoes multiple 100% inspections.In addition, the work order underwent functional product verification testing on a sampling basis as a requirement for lot release.These inspections and tests during the manufacturing process support that it is unlikely that a non-conformance contributed to the reported event.In this case, review of available information suggests that patient factors (calcification) contributed to the balloon burst.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.The instructions for use (ifu)/training manuals were reviewed for guidance/instruction involving the commander delivery system usage.Based on the review of the ifu/training manuals, no deficiencies were identified.Therefore, no product risk assessment no corrective or preventative action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMMANDER DELIVERY SYSTEM
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key17882853
MDR Text Key325082473
Report Number2015691-2023-16585
Device Sequence Number1
Product Code NPT
UDI-Device Identifier00690103207835
UDI-Public(01)00690103207835(17)250123(10)64849909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9750CM26A
Device Lot Number64849909
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/05/2023
Supplement Dates Manufacturer Received11/16/2023
Supplement Dates FDA Received11/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-