• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS SURESIGNS VS4 NBP, SPO2

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS SURESIGNS VS4 NBP, SPO2 Back to Search Results
Model Number 863283
Device Problem No Audible Prompt/Feedback (2282)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/26/2023
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) went onsite and observed the audio alarm error.The fse performed an audio test which cleared the error.The audio worked correctly, and the alarm was no longer present.The device is back in use.The reported problem was confirmed.The device was operational after the audio test was completed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Event Description
Philips received a complaint on the suresigns vs4 nbp, spo2 indicating that the machine was showing a loudspeaker failure and audio failure.Unable to hear alarms.The device was not in use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESIGNS VS4 NBP, SPO2
Type of Device
SURESIGNS VS4 NBP, SPO2
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key17884760
MDR Text Key325067092
Report Number1218950-2023-00732
Device Sequence Number1
Product Code DSJ
UDI-Device Identifier00884838087095
UDI-Public00884838087095
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K163649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number863283
Device Catalogue Number863283
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/26/2023
Initial Date FDA Received10/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/28/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-