• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. MODULAR TIBIA KEEL BROACH SIZE 1 / 2; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. MODULAR TIBIA KEEL BROACH SIZE 1 / 2; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem Foreign Body In Patient (2687)
Event Date 03/31/2023
Event Type  Injury  
Event Description
It was reported that during a knee procedure, part of the impactor fractured and was retained by the patient.No additional patient consequences were reported.
 
Manufacturer Narrative
(b)(4).Proposed component code: mechanical (g04) - impactor.Fracture of the instrument was confirmed by review of photographs.Visual examination of the provided pictures identified a fractured broach.Patient retention of a foreign body could not be confirmed.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULAR TIBIA KEEL BROACH SIZE 1 / 2
Type of Device
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17885432
MDR Text Key325078181
Report Number0001822565-2023-02725
Device Sequence Number1
Product Code JWH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K023528
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number627902102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/23/2023
Initial Date FDA Received10/06/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-