Manufacturer's investigation conclusion: the reported event of a malformation on the motor cable near the connector was confirmed via evaluation.The centrimag motor, ous (s/n: (b)(6)) was sent to the european distribution center (edc) for preventative maintenance.Visual inspection revealed a bulge at the motor cable connector.The motor was forwarded to the product performance engineering (ppe) group for further investigation.The forwarded motor underwent resistance and insulation testing and passed with no issues, even when manipulating the cable.The unit was connected to test equipment and a mock circulatory loop and was able to operate the system as intended for extended operation.The motor cable jacket was opened where damage was observed.No damage to the underlying wires or loose connections were found.No alarm was associated with the reported event.A root cause for the reported event was unable to be conclusively determined through this analysis.The device history records were reviewed for centrimag motor and the motor was found to pass all manufacturing and qa specifications.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." centrimag motor ifu instructs the user to inspect the centrimag motor, cable, console connector, and locking mechanism for any damage prior to use.If any component is damaged, do not use the centrimag motor.This document states that if the unit fails to operate according to the motor specifications or a console diagnostic error indicates a centrimag motor malfunction, it should be returned.Additionally, this document instructs the user to always have a spare centrimag motor and back-up equipment available.No further information was provided.The manufacturer is closing the file on this event.
|