• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, OUS; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, OUS; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 201-10002
Device Problems Mechanical Problem (1384); Electrical Power Problem (2925)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/18/2023
Event Type  malfunction  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
 
Event Description
It was reported that during testing of the centrimag motor, a malformation was found on the cable at the connector.When moving the cable lower than 0.7o value was being observed measuring between a1+ and a1-.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of a malformation on the motor cable near the connector was confirmed via evaluation.The centrimag motor, ous (s/n: (b)(6)) was sent to the european distribution center (edc) for preventative maintenance.Visual inspection revealed a bulge at the motor cable connector.The motor was forwarded to the product performance engineering (ppe) group for further investigation.The forwarded motor underwent resistance and insulation testing and passed with no issues, even when manipulating the cable.The unit was connected to test equipment and a mock circulatory loop and was able to operate the system as intended for extended operation.The motor cable jacket was opened where damage was observed.No damage to the underlying wires or loose connections were found.No alarm was associated with the reported event.A root cause for the reported event was unable to be conclusively determined through this analysis.The device history records were reviewed for centrimag motor and the motor was found to pass all manufacturing and qa specifications.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." centrimag motor ifu instructs the user to inspect the centrimag motor, cable, console connector, and locking mechanism for any damage prior to use.If any component is damaged, do not use the centrimag motor.This document states that if the unit fails to operate according to the motor specifications or a console diagnostic error indicates a centrimag motor malfunction, it should be returned.Additionally, this document instructs the user to always have a spare centrimag motor and back-up equipment available.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG MOTOR, OUS
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key17886014
MDR Text Key325086877
Report Number3003306248-2023-06980
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K020271
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number201-10002
Device Catalogue Number201-10002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/18/2023
Initial Date FDA Received10/06/2023
Supplement Dates Manufacturer Received01/30/2024
Supplement Dates FDA Received02/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-