The issue was resolved by hopistal's biomed replacing the t-connection fittings and hoses.A device service history review has been performed and identified that the unit was manufactured in 2022 and no other similar events have been reported.Based on the available elements, it is reasonable to assume that the wear and deterioration of the connection resulted in the reported event.Based on the information received, water leak from the device is negligible ("slow drip") and not massive.Thus, the issue is not likely to result in serious injury.The event has been re-assesed as not reportable.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|