Trackwise # (b)(4).Updated section: b4, d10, g4, g7, h2, h3, h6, h10.The device was returned to the factory for evaluation on 10/06/2023.An investigation was conducted on 10/10/2023.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.The power supply was observed to be intact.An electrical evaluation was conducted.A reference power cord was plugged into the power supply and the device was switched on.The green lights on the power supply did not turn on.A pre-cautery test was performed per the instruction for use (ifu) with a reference adapter, hp2, cable, and power supply vh-3010 at level 3.0.The device did not pass the pre-cautery test.Based on the returned condition of the device, the reported failure "failure to deliver energy" was confirmed.The reported device is an oem device.The certificate of conformance was reviewed for the serial # (b)(6).The vendor certifies that this device serial # conforms to all applicable product specifications and there were no non-conformances identified for the manufacturing batch.
|