Model Number 866060
Device Problem Device Fell (4014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/11/2023
Event Type  malfunction  
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Event Description
The customer reported that the device fell.In addition, it was reported that the mounting was defective.It is unknown if the device was in use at time of event, and there was no adverse event reported.
 
Manufacturer Narrative
Philips received a complaint on the intellivue mx400 patient monitor indicating that the device fell significantly.A good faith effort (gfe) was performed to clarify the cause of the device fall, but no additional details were provided.The device was sent to philips bench repair.A philips bench repair technician (brt) evaluated the device and found that the touch bezel, handle, and battery door were damaged.The brt also indicated that the small parts kit and screw kits were needed as well.Based on the information available and the testing conducted, the cause of the reported problem is unknown.The reported problem was not confirmed.Based on the information available and results of additional analysis, no further action is necessary at this time.The device was operational after replacing the handle, touch bezel, battery door, small parts kit, and screw kit.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MX400 PATIENT MONITOR
Type of Device
INTELLIVUE MX400 PATIENT MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key17886328
MDR Text Key325090861
Report Number9610816-2023-00508
Device Sequence Number1
Product Code MHX
UDI-Device Identifier00884838038752
UDI-Public00884838038752
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K130849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866060
Device Catalogue Number866060
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/11/2023
Initial Date FDA Received10/06/2023
Supplement Dates Manufacturer Received01/08/2024
Supplement Dates FDA Received02/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-