• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977762
Device Problem Device Damaged by Another Device (2915)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2023
Event Type  malfunction  
Event Description
A stokist reported that during surgery the cartridge chewed up the lens.Additional information was requested, received and stated there was no patient harm.
 
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The first photo showed the lens carton endcap with the serial number.The lens was in from of the carton, posterior surface up.Viscoelastic and possible blood was observed dried on the lens.The haptics were intact.No torn "chewed" areas were observed.The second photo was the bottom view of a company cartridge.The view was from the loading area to mid-nozzle.The tip was not visible.Unable to observe/determine viscoelastic amount.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17892593
MDR Text Key325810477
Report Number1119421-2023-01719
Device Sequence Number1
Product Code MSS
UDI-Device Identifier00380659777622
UDI-Public00380659777622
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K112977
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065977762
Device Lot Number15617184
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/14/2023
Initial Date FDA Received10/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/20/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ACRYSOF IQ NATURAL SINGLEPIECE IOL.
Patient Age65 YR
Patient SexFemale
-
-