MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3013-55
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/02/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported during a routine check, the cs100 intra-aortic balloon pump(iabp) screen flickers when turned on.There was no patient involvement.
 
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the unit and found that when turning on the unit screen flashes all the time.The fse removed the video receiver board and connector cable to clean and put it back in again.After cleaning the connector area that is connected to all pcb, video receiver, it was found that the iabp unit can be used normally.The unit was then in good working condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key17895512
MDR Text Key325964823
Report Number2249723-2023-04361
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107295
UDI-Public10607567107295
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3013-55
Device Catalogue Number0998-00-3013-55
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/02/2023
Initial Date FDA Received10/09/2023
Supplement Dates Manufacturer Received11/14/2023
Supplement Dates FDA Received11/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/13/2006
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-