Model Number DB-1416
Device Problems Premature Discharge of Battery (1057); Battery Problem (2885)
Patient Problems Pain (1994); Shaking/Tremors (2515)
Event Date 09/08/2023
Event Type  Injury  
Event Description
It was reported that the deep brain stimulation (dbs) patient experienced moderate to severe pain due to a return of their parkinsons disease symptoms.It was confirmed that the patient received the end of life (eol) message on their remote control resulting in a loss of stimulation.Additionally, it was confirmed that the patient had used high rate stimulation settings prior to the loss of therapy.Therefore, the patient underwent an implantable pulse generator (ipg) revision procedure during which the ipg was explanted and replaced.The patient has fully recovered and is doing well postoperatively.The explanted ipg will not be returned as it was retained by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERCISE GENUS
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key17901605
MDR Text Key325234020
Report Number3006630150-2023-06125
Device Sequence Number1
Product Code NHL
UDI-Device Identifier08714729985020
UDI-Public08714729985020
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/08/2023
Device Model NumberDB-1416
Device Catalogue NumberDB-1416
Device Lot Number207081
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/13/2023
Initial Date FDA Received10/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age74 YR
Patient SexMale