• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PERFUSION SYSTEMS EOPA ARTERIAL CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PERFUSION SYSTEMS EOPA ARTERIAL CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 77422
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/07/2023
Event Type  malfunction  
Event Description
Medtronic received information that prior to use of an eopa arterial cannula, it was reported that upon removing the cannula from the packaging the vented plug snapped very easily where the vented plug exits the introducer hub.The device was replaced.There was no patient involvement, so no adverse effect occurred.The customer stated that the experienced user of cannula who handled this device suggested that this was very unusual as very little pressure was applied, they handled the cannula as normal.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Device evaluation summary: visual inspection shows the porous plug is broken.Reason for return was confirmed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EOPA ARTERIAL CANNULA
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17903770
MDR Text Key325254376
Report Number2184009-2023-00973
Device Sequence Number1
Product Code DWF
UDI-Device Identifier00763000135638
UDI-Public00763000135638
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K031037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number77422
Device Catalogue Number77422
Device Lot Number2022100560
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/14/2023
Initial Date FDA Received10/10/2023
Supplement Dates Manufacturer Received09/14/2023
Supplement Dates FDA Received11/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-