BOSTON SCIENTIFIC CORPORATION ROTAPRO; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 39467-150
Device Problems Failure to Advance (2524); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/19/2023
Event Type  malfunction  
Event Description
It was reported that unstable speed occurred.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified artery.A 1.50mm rotapro was selected for percutaneous coronary intervention (pci) procedure.During the procedure, it was noted that the burr was unable to cross into the 7f non-boston scientific guide catheter used while on dynaglide mode.After four ablations, it was noted that the rotation speed was unstable.The set rotation speed was at 190,000rpm while the maximum speed observed was at 185,000-195,000rpm during procedure.The procedure was completed with another of the same device.No patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAPRO
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17904574
MDR Text Key325279642
Report Number2124215-2023-51746
Device Sequence Number1
Product Code MCX
UDI-Device Identifier08714729893356
UDI-Public8714729893356
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number39467-150
Device Catalogue Number39467-150
Device Lot Number0031099312
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/19/2023
Initial Date FDA Received10/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/22/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GUIDE CATHETER: ASAHI HYPERION 7F SAL 1.0
-
-