U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW CVC SET: 3-LUMEN 7 FR X 20 CM; CATHETER PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW CVC SET: 3-LUMEN 7 FR X 20 CM; CATHETER PERCUTANEOUS Back to Search Results
Catalog Number CV-15703
Device Problems Unraveled Material (1664); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/07/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported the spring wire guide unraveled when taking it out of the introducer needle.No patient harm was reported.The patient's condition is reported as fine.
 
Manufacturer Narrative
Qn# (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event Description
It was reported the spring wire guide unraveled when taking it out of the introducer needle.No patient harm was reported.The patient's condition is reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 3-LUMEN 7 FR X 20 CM
Type of Device
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
mariah mackinnon
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key17905277
MDR Text Key325293775
Report Number9680794-2023-00764
Device Sequence Number1
Product Code DQY
UDI-Device Identifier10801902105339
UDI-Public10801902105339
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K862056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberCV-15703
Device Lot Number14F22G0209
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/13/2023
Initial Date FDA Received10/10/2023
Supplement Dates Manufacturer Received10/27/2023
Supplement Dates FDA Received10/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/28/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NOT REPORTED.; NOT REPORTED.
-
-