C.R. BARD, INC. (BASD) -3006260740 M.R.I. HARD BASE IMPLANTABLE PORT, HICKMAN SINGLE-LUMEN, 6.6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 0604580CE |
Device Problems
Fracture (1260); Difficult to Remove (1528); Material Separation (1562); Stretched (1601); Failure to Advance (2524); Deformation Due to Compressive Stress (2889); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/08/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during the port placement procedure, the guide allegedly stuck in the puncture needle.It was further reported that the assembly was removed by forcing which deformed the guide.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the m.R.I.Low-profile implantable port, hickman single-lumen, 6.6f products that are cleared in the us.The pro code and 510 k number for the m.R.I.Low-profile implantable port, hickman single-lumen, 6.6f products are identified in d2 and g4.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiry date: 07/2026).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the m.R.I.Low-profile implantable port, hickman single-lumen, 6.6f products that are cleared in the us.The pro code and 510 k number for the m.R.I.Low-profile implantable port, hickman single-lumen, 6.6f products are identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one introducer needle and a guidewire with a straightener were returned for evaluation.Functional, gross visual and microscopic visual evaluations were performed.The guidewire was noted to be uncoiled throughout.The flat inner core wire was noted to have a complete break within the guidewire.Therefor based on investigation reported deformation and identified the fracture, material separation and stretched issue can be confirmed.However, the investigation is inconclusive for the reported physical sticking, failure to advance and difficult to remove issues, as the exact circumstances at the time of the reported event was unknown.Based on the measurements recorded during sample evaluation and applicable drawings, no measurements recorded could be confirmed to be out of tolerance.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiry date: 07/2026), h6 (device).H11: h6 (method, result, conclusion).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during the port placement procedure, the guide allegedly stuck in the puncture needle.It was further reported that the assembly was removed by forcing which deformed the guide.There was no reported patient injury.
|
|