• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA APOLLO; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA APOLLO; ANESTHESIA UNITS Back to Search Results
Catalog Number 8606500
Device Problems Gas Output Problem (1266); Failure to Deliver (2338); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/30/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation was performed based on the log analysis whereby the reported ventilator failure could be confirmed.No indications for a device malfunction being root cause for the ventilator failure were found.The reported symptom was caused by an overpressure at the patient end of the circuit leading to a pressure peak.Most likely, the patient was coughing against the ventilator at that time explaining the pressure peak.To prevent from damages, the system is designed to shut down automatic ventilation and to alert the user to this condition by means of a corresponding alarm.Manual ventilation and the monitoring functions remain available to the full extent.The device reacted as specified for the detected situation.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.H3 other text : on-going.
 
Event Description
It was reported that a ventilator failure occurred during a case.No injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APOLLO
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
Manufacturer Contact
moislinger allee 53-55
lübeck 23542
4518822868
MDR Report Key17910912
MDR Text Key325344389
Report Number9611500-2023-00358
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier04048675229674
UDI-Public(01)04048675229674(11)161102(17)170327(93)8606500-63
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 09/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8606500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/14/2023
Initial Date FDA Received10/11/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/31/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-