• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLFLEX DUODENAL; STENT,METALLIC,EXPANDABLE,DUODENAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLFLEX DUODENAL; STENT,METALLIC,EXPANDABLE,DUODENAL Back to Search Results
Model Number M00565010
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/18/2023
Event Type  Injury  
Event Description
It was reported to boston scientific corporation that a wallflex enteral duodenal stent was to be implanted in the duodenum to treat a 2cm malignant tumor during an endoscopic retrograde cholangiopancreatography (ercp) with stent placement procedure performed on (b)(6) 2023.The patient's anatomy was tortuous.During the procedure, the stent was deployed; however, it was noticed that the stent was kinked.Subsequently, the stent was removed using foreign body forceps.The procedure was completed with another wallflex enteral duodenal stent.In addition, it was noted that action taken by the physician to resolve the event also included observation.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Block e1: the initial reporter's facility name is (b)(6).Block h6: imdrf device code a0406 captures the reportable event of stent material deformation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX DUODENAL
Type of Device
STENT,METALLIC,EXPANDABLE,DUODENAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17912479
MDR Text Key325393983
Report Number3005099803-2023-05428
Device Sequence Number1
Product Code MUM
UDI-Device Identifier08714729456483
UDI-Public08714729456483
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K062750
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00565010
Device Catalogue Number6501
Device Lot Number0030596266
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/19/2023
Initial Date FDA Received10/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age61 YR
Patient SexMale
Patient Weight72 KG
-
-