• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION INC AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION INC AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems No Display/Image (1183); Display or Visual Feedback Problem (1184); Use of Device Problem (1670); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2023
Event Type  malfunction  
Manufacturer Narrative
The customer reported complaint that the autopulse platform (sn (b)(6)) makes a clicking sounds upon power up, the display screen showed just a single line through it, and the platform is unable to retract lifebands was confirmed during functional testing.The root cause for the reported complaint was a failed processor board, likely attributed to a failed component.During visual inspection, unrelated to the reported complaint, a sticky driveshaft was observed.The root cause was the sticky driveshaft clutch area, which is usually caused by sharp edges from all 12-hex edges of the armature plate, or due to burrs on the clutch rotor's surface likely attributed to normal wear and tear.The sticky clutch's impact was not severe enough to make the platform non-functional.The clutch plate was deburred to remedy the problem.The data archive could not be retrieved due to the failed processor board.The autopulse platform failed functional testing.The platform displayed a blank white screen and a continuous clicking noise was observed at power up; thus, confirming the reported complaint.Due to this, the platform would also be unable to retract the lifeband.The processor pca assembly was replaced to remedy the problem.Subsequently, the device was tested with an instrumented manikin and 95% patient large resuscitation test fixture (lrtf) for approximately 15 minutes each with no issue.After service, the autopulse platform passed the run-in test without any fault or error.The autopulse platform passed the final testing without any fault or error.Historical complaints were reviewed for service information related to the reported complaint, and there was no previous history of complaints for autopulse platform with sn (b)(6).
 
Event Description
During shift check, the autopulse platform (sn (b)(6)) makes a clicking sounds upon power up and the display screen showed just a single line through it.The customer tested the platform with three different autopulse li-ion batteries and two different lifebands with the same results.The platform is unable to retract the lifebands.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION INC
2000 ringwood ave
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION INC
2000 ringwood ave
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave
san jose, CA 95131
4084192922
MDR Report Key17913406
MDR Text Key325405260
Report Number3010617000-2023-00855
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/13/2023
Initial Date FDA Received10/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/09/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-