Catalog Number 35700BAX2
Device Problems False Alarm (1013); Failure to Power Up (1476); Failure to Reset (1532)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a spectrum pump alarmed constant downstream occlusion and did not restart.The event occurred during testing in the biomedical service department.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Correction: h6 device problem code of a070803 corrected to a160501 additional information: d9, h3, h6, h10 h10: the device was received for evaluation.During functional testing, the reported problem "constant downstream too low & does not restart unable to remedy" was not reproduced during downstream occlusion testing.A service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.The reported condition was not verified.No pump correction is required.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRUM INFUSION PUMP
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEDINA
711 park ave
medina NY 14103
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17914222
MDR Text Key325412682
Report Number1314492-2023-03724
Device Sequence Number1
Product Code FRN
UDI-Device Identifier00085412498683
UDI-Public(01)00085412498683
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K133801
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number35700BAX2
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/19/2023
Initial Date FDA Received10/11/2023
Supplement Dates Manufacturer Received01/08/2024
Supplement Dates FDA Received01/10/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
-
-