ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; Permanent defibrillator electrodes Back to Search Results
Model Number 7122Q/58
Device Problems Failure to Capture (1081); Over-Sensing (1438); Failure to Sense (1559); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/16/2023
Event Type  Injury  
Event Description
It was reported that the patient presented during clinical follow-up.Upon interrogation, it was noted that the right ventricular lead exhibited far p wave oversensing which indicated possible dislodgement.Dislodgement was confirmed on (b)(6) 2023 in clinic and it was also noted that the right ventricular lead failed to capture and failed to sense.The reported lead was explanted and replaced on (b)(6) 2023.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
Permanent defibrillator electrodes
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17914224
MDR Text Key325412714
Report Number2017865-2023-48370
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734503174
UDI-Public05414734503174
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7122Q/58
Device Lot NumberA000124499
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/20/2023
Initial Date FDA Received10/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TENDRIL LEAD; TENDRIL LEAD; UNIFY ASSURA
Patient Outcome(s) Required Intervention;
Patient Age74 YR
Patient SexMale
-
-