Model Number CYF-V2
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/25/2023
Event Type  malfunction  
Event Description
The customer reported to olympus, the bending rubber of the visera cysto-nephro videoscope was broken and the metal was sticking out.The scope cap was not on during the the sterilization period.The issue was found during reprocessing.There were no reports of patient harm.
 
Manufacturer Narrative
The device was returned to olympus for evaluation, however, no information has been provided at the moment.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause of the reported ¿bending rubber is broken/torn/ruptured and the metal is sticking out¿ and ¿device washed with sterilization cap on¿ issues could not be determined, as the device has not been returned to olympus for evaluation, however, the reported issues were likely the result of user mishandling.The event can be prevented by following the instructions for use which state: olympus ltd type v2 - chapter 6 compatible reprocessing methods and chemical agents 6.6 ethylene oxide gas sterilization caution attach the sterilization cap to the endoscope before ethylene oxide gas sterilization.If the sterilization cap is not attached to the endoscope during ethylene oxide gas sterilization, the air inside the endoscope will expand and could rupture the bending section cover and/or damage the angulation mechanism (see figure 6.2).Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17917117
MDR Text Key326260694
Report Number3002808148-2023-11153
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170411199
UDI-Public04953170411199
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K221683
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-V2
Was Device Available for Evaluation? No
Date Returned to Manufacturer10/02/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/02/2023
Initial Date FDA Received10/11/2023
Supplement Dates Manufacturer Received11/24/2023
Supplement Dates FDA Received12/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/16/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-