Model Number 15 |
Device Problems
Incorrect Interpretation of Signal (1543); Failure to Sense (1559)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/12/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Stryker contacted the customer to request additional information on the patient.No response has been received from the customer.Stryker evaluated the customer's device and was unable to duplicate the reported issue.Proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer for use.
|
|
Event Description
|
A customer contacted stryker to report that their device did not detect the patient being connected and would not advise shock for a rhythm they believed to be shockable.The customer put the device in manual mode and was successfully able to continue patient care.There were no reports of adverse effects to the patient as a result of the reported event.
|
|
Manufacturer Narrative
|
The electronic patient records were downloaded and reviewed.The connect electrodes message was able to be confirmed and is present and it was observed that the rhythm is consistent with ventricular fibrillation.The analysis shows multiple shock advised followed by a leads off event.This is indicative of poor connection to the patient.The algorithm properly analyzed the shockable rhythm but did not provide shock due to poor leads connection.This is the root cause of the reported issue.Further root cause could not be determined.
|
|
Event Description
|
A customer contacted stryker to report that their device did not detect the patient being connected and would not advise shock for a rhythm they believed to be shockable.The customer put the device in manual mode and was successfully able to continue patient care.There were no reports of adverse effects to the patient as a result of the reported event.
|
|
Search Alerts/Recalls
|