During device evaluation at olympus, it was found the complaint was confirmed and the image went out due to a faulty scope socket.The locking mechanism and scope detection slide were not functioning, which caused the front panel to intermittently blink.Evaluation also found the housing had minor cosmetic wear and tear and a non-olympus lamp had 50 or more hours.This event is under investigation.A supplemental report will be submitted upon receiving additional information.
|
This supplemental report is being submitted to provide the correction of the initial mdr and the legal manufacturer's final investigation.Additional information: h6: (add a medical device problem code and investigation code).A review of the device history record found no deviations, that could have caused or contributed to the reported issue.For the image problem, when scope is manipulated.Issue a definitive root cause was not identified.However, based on the results of the investigation, the probable cause of the malfunction would likely, be a faulty connector (contact failure or the like).For the front panel blinking constantly, due to the locking mechanism and scope detection slide failure.A definitive root cause was not identified.However, based on the results of the investigation, the probable cause of the malfunction would likely, be the locking mechanism and the scope detection slide did not work.The scope detection function did not work properly.The event can be prevented, by following the instructions for use which state: [warning] non-olympus equipment can cause instability of the automatic brightness adjustment.Olympus will continue to monitor field performance for this device.
|