Product complaint # (b)(4).H6 component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary: no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
It was reported that the patient came to doctor's office on tuesday (b)(6) due to a sudden onset of redness, swelling and drainage of her left hip at her surgical incision.Patient had a total hip done by doctor on (b)(6) 2023.Doctor determined the patient hip is infected.Surgeon had admitted the patient and scheduled an i&d of the pt's left hip.During surgery it was noted that the tissues and fluid that came out of the hip during debridement was consistent with infection.Doctor removed the poly liner and femoral head so that those areas could be cleaned.It was noted that all implants were well fixed and no deficiencies were seen.Doctor cleaned and derided the wound and then a new poly liner and femoral head was placed.The wound was then closed.No other information is available at this time.There was no surgical delay.Doi: aug 21, 2023.Dor: (b)(6) 2023.Affected side: left hip.
|