Model Number LEV2517-MVE |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Hyperplasia (1906); Vertigo (2134); Stenosis (2263)
|
Event Type
Injury
|
Event Description
|
As reported in the sealant study, on (b)(6) 2023, the patient was treated for a saccular bifurcation aneurysm, located on the basiliar tip.Measurement of the aneurysm: h-w-n (3.3x 3.9 x 3.2).Parent artery (1.8 x 2.6 x 109).Pre procedure neurological evaluation: mrs score 0 - nihss score 0.The aneurysm never ruptured and not previously treated.The treatment consisted of using a lvis evo lev2517 stent.Two coils were used during the procedure under jailed-catheter technique.The parent artery at the end of the procedure remains without stenosis.On (b)(6) 2023 the patient was discharged from the hospital with a neurological evaluation; mrs 0.The antiplatelet test resistance was not performed.In (b)(6) 2023; three months post-stent implantation (exact date unknown), the patient experienced vertigo and vertical diplopia 2 times.Additionally, hyperplasia and intra-stent 50% stenosis occurred.The action taken in regards of this ae is a drug treatment with ¿reintroduction of clopidogrel 75mg per day¿, head mri and cerebral angiography.The outcome was resolved without sequelae on (b)(6) 2023.It was noted the event is most likely related to the lvis evo device and not the index endovascular procedure nor the other devices used.The event is also not related to the study disease condition and not concurrent disease condition or treatment.This complaint addresses the vertigo episode and vertical diplopia as it is defined as related to the study device.In addition, the complaint addresses the neurological event: hyperplasia and intra-stent 50% stenosis.
|
|
Manufacturer Narrative
|
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was implanted in the patient and not returned to the manufacturer for evaluation.Procedural or medical imaging was not provided.The alleged incident, issues and stenosis as described could not be confirmed at this time.If imaging is received at a later date, an investigation will be performed and a supplemental mdr will be submitted.H3 other text : remains implanted.
|
|
Event Description
|
No additional information was provided, please see h6 and h10.
|
|
Manufacturer Narrative
|
Items returned: - n/a visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation, nor were any images of the device provided in place of a device return.Procedure and medical imaging was not provided for this investigation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): potential complications possible complications include but are not limited to the following: ¿ hematoma at the puncture site ¿ perforation or dissection of the vessel(s) ¿ intravascular spasm ¿ hemorrhaging ¿ rupture or perforation of aneurysm ¿ coil herniation ¿ device migration ¿ neurologic insufficiencies including stroke and death ¿ ischemia ¿ vascular occlusion ¿ vessel stenosis ¿ incomplete aneurysm occlusion ¿ pseudoaneurysm formation ¿ distal embolization ¿ headache ¿ infection ¿ reaction to contrast agents including severe allergic reactions and renal failure warnings should unusual resistance be felt at any time during access or removal, the introducer/guide catheter/microcatheter and lvis evo device should be removed as a single unit.Applying excessive force during delivery or retrieval of the lvis evo device can potentially result in loss or damage to the device and delivery components.It is imperative to use the lvis evo device with compatible microcatheters.If repeated friction is encountered during lvis evo device delivery, verify microcatheter is not kinked or in extremely tortuous anatomy.Confirm that the microcatheter does not ovalize.Confirm that there is adequate sterile flush solution.Do not reposition the lvis evo device in the parent vessel without fully retrieving the device.The lvis evo device must be retrieved into the microcatheter and re-deployed at the desired target location or removed completely from the patient.Precautions exercise caution when crossing the deployed/detached lvis evo device with adjunctive devices such as guidewires, catheters, microcatheters or balloon catheters to avoid disrupting the device geometry and device placement.Directions for use 15.Advance the delivery wire to transfer the lvis evo device from within the introducer into the microcatheter.Warning: do not torque the delivery wire while advancing or retracting the lvis evo device.A torque device should not be used.16.Continue advancing the delivery wire into the microcatheter until the proximal tip of the delivery wire enters the introducer.Loosen the rhv locking ring, remove the introducer, and set it aside.Warning: do not apply undue force.If resistance is encountered at any point during lvis evo device delivery or manipulation, withdraw the unit and select a new lvis evo device.18.Position the lvis evo device for deployment by aligning the lvis evo implant distal radiopaque end markers approximately 7 mm or adequate length past the aneurysm neck.Note: a proper push/pull technique, encompassing sufficient delivery wire push force, in addition to an opposing microcatheter withdrawal force, will facilitate properly deploying the lvis evo device to achieve full expansion and good vessel apposition.Note: slowly advancing the lvis evo device while adjusting the microcatheter position will ensure accurate deployment.Maintain simultaneous control of the lvis evo device and microcatheter in order to position and expand the device at the proper location.Caution: using a rapid microcatheter withdrawal technique to deploy the lvis evo device is not recommended and may result in device elongation.19.If lvis evo device positioning is not satisfactory, the lvis evo device may be recaptured and repositioned if it is not fully deployed.The lvis evo device may be recaptured until the point where the proximal end of the lvis evo device markers is aligned 3 mm proximally with the microcatheter distal marker band.Caution: if resistance is felt while recapturing the lvis evo device, do not continue to recapture the device.Withdraw the microcatheter slightly to unsheath the lvis evo device (without exceeding the recapture limit), and then attempt to recapture the lvis evo device.Caution: the lvis evo device must not be re-deployed more than three times.Note: the lvis evo device delivery wire should not be utilized as a guidewire.Do not torque the lvis evo device.A torque device should not be used.20.If lvis evo device positioning is satisfactory, carefully retract the microcatheter and advance the delivery wire together, to allow the lvis evo device to deploy across the neck of the aneurysm.Ensure the device proximal radiopaque end markers are approximately 7 mm or adequate length proximal to the aneurysm neck to ensure an adequate landing zone.The lvis evo device will expand and total length may foreshorten up to 60% from its undeployed length (refer to table 1) as it exits the microcatheter.Ensure microcatheter is retracted and clear from the proximal flared ends.Note: visualize and refer to the implant radiopaque end markers to maintain adequate implant length, approximately 7 mm or adequate length on each side of the aneurysm neck or target location to ensure appropriate neck coverage.Warning: do not detach the lvis evo device if it is not properly positioned in the parent vessel.Observe the delivery wire distal tip to assure it remains within the desired location of the parent vessel.21.Prior to removing the delivery wire and if necessary, carefully position the microcatheter distal to the lvis evo device to maintain access through the lvis evo device.Remove and discard the delivery wire.Warning: the lvis evo device delivery wire should not be utilized as a guidewire.Do not torque the lvis evo device.A torque device should not be used.23.Use the guidewire and microcatheter to access the aneurysm through the lvis evo device cells.Warning: observe lvis evo device marker position during placement of the microcatheter into the aneurysm to ensure that the lvis evo device does not migrate or dislodge from its deployed position.Note: access to the aneurysm may be facilitated by the use of a microcatheter that has been shaped.25.Warning: observe lvis evo device marker position during the coiling procedure to ensure that the device does not migrate from its deployed position.After placing the last coil, verify that the lvis evo device has remained patent and properly positioned.Advance a guidewire, if necessary, to the microcatheter tip and carefully remove the microcatheter.Note: a microcatheter may be positioned into the aneurysm sac prior to delivery of the lvis evo device.The microcatheter will be supported by the lvis evo device during delivery of embolic coiling.After completing the coiling, the coiling microcatheter should be carefully removed to avoid dislodging the lvis evo device.27.Caution: carefully watch the lvis evo device distal and proximal markers when passing through the deployed lvis evo device with embolic coiling microcatheters to avoid displacing the lvis evo device.The physical device was not available for evaluation to determine if a condition existed that would have caused or contributed to event.Supplemental imaging was also unavailable for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.This information may be updated if additional information is provided at a later date.
|
|
Manufacturer Narrative
|
Procedure/medical information review: two dicom files are supplied and were reviewed.Left vertebral and left ica angiogram, (b)(6) 2023: left vertebral: ap and lateral dsa images, subtracted and unsubtracted, with and without contrast.An lvis evo spans from the distal basilar artery to the right p1/p2 junction; it appears well-opened.There may be very mild distal fish mouthing.There are coils in a small basilar tip aneurysm.There is a 59% in-stent stenosis of the entire stent, as shown on an ap image where measurements were taken at the level of the mid-p1: residual lumen: 0.99 mm, stent itself: 2.40mm.There is no flow limitation.Left ica: ap and lateral dsa images, subtracted and unsubtracted, with and without contrast.Normal angiogram.Cerebral mri/mra, (b)(6) 2023: multiple sequences were obtained.The mra redemonstrates the in-stent stenosis; there is no evidence of ischemic infarcts.There is a small susceptibility artifact in the mid right cerebellum, of unknown origin.No images of the embolization procedure ((b)(6)2023) or of the first follow-up (at the time of symptoms, in (b)(6) 2023) are supplied.These images do not explain why the in-stent stenosis occurred.Items returned: - n/a.Visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): potential complications possible complications include but are not limited to the following: ¿ hematoma at the puncture site.¿ perforation or dissection of the vessel(s).¿ intravascular spasm.¿ hemorrhaging.¿ rupture or perforation of aneurysm.¿ coil herniation.¿ device migration.¿ neurologic insufficiencies including stroke and death.¿ ischemia.¿ vascular occlusion.¿ vessel stenosis.¿ incomplete aneurysm occlusion.¿ pseudoaneurysm formation.¿ distal embolization.¿ headache.¿ infection.¿ reaction to contrast agents including severe allergic reactions and renal failure.Warnings should unusual resistance be felt at any time during access or removal, the introducer/guide catheter/microcatheter and lvis evo device should be removed as a single unit.Applying excessive force during delivery or retrieval of the lvis evo device can potentially result in loss or damage to the device and delivery components.It is imperative to use the lvis evo device with compatible microcatheters.If repeated friction is encountered during lvis evo device delivery, verify microcatheter is not kinked or in extremely tortuous anatomy.Confirm that the microcatheter does not ovalize.Confirm that there is adequate sterile flush solution.Do not reposition the lvis evo device in the parent vessel without fully retrieving the device.The lvis evo device must be retrieved into the microcatheter and re-deployed at the desired target location or removed completely from the patient.Precautions exercise caution when crossing the deployed/detached lvis evo device with adjunctive devices such as guidewires, catheters, microcatheters or balloon catheters to avoid disrupting the device geometry and device placement.Directions for use 15.Advance the delivery wire to transfer the lvis evo device from within the introducer into the microcatheter.Warning: do not torque the delivery wire while advancing or retracting the lvis evo device.A torque device should not be used.16.Continue advancing the delivery wire into the microcatheter until the proximal tip of the delivery wire enters the introducer.Loosen the rhv locking ring, remove the introducer, and set it aside.Warning: do not apply undue force.If resistance is encountered at any point during lvis evo device delivery or manipulation, withdraw the unit and select a new lvis evo device.18.Position the lvis evo device for deployment by aligning the lvis evo implant distal radiopaque end markers approximately 7 mm or adequate length past the aneurysm neck.Note: a proper push/pull technique, encompassing sufficient delivery wire push force, in addition to an opposing microcatheter withdrawal force, will facilitate properly deploying the lvis evo device to achieve full expansion and good vessel apposition.Note: slowly advancing the lvis evo device while adjusting the microcatheter position will ensure accurate deployment.Maintain simultaneous control of the lvis evo device and microcatheter in order to position and expand the device at the proper location.Caution: using a rapid microcatheter withdrawal technique to deploy the lvis evo device is not recommended and may result in device elongation.19.If lvis evo device positioning is not satisfactory, the lvis evo device may be recaptured and repositioned if it is not fully deployed.The lvis evo device may be recaptured until the point where the proximal end of the lvis evo device markers is aligned 3 mm proximally with the microcatheter distal marker band.Caution: if resistance is felt while recapturing the lvis evo device, do not continue to recapture the device.Withdraw the microcatheter slightly to unsheath the lvis evo device (without exceeding the recapture limit), and then attempt to recapture the lvis evo device.Caution: the lvis evo device must not be re-deployed more than three times.Note: the lvis evo device delivery wire should not be utilized as a guidewire.Do not torque the lvis evo device.A torque device should not be used.20.If lvis evo device positioning is satisfactory, carefully retract the microcatheter and advance the delivery wire together, to allow the lvis evo device to deploy across the neck of the aneurysm.Ensure the device proximal radiopaque end markers are approximately 7 mm or adequate length proximal to the aneurysm neck to ensure an adequate landing zone.The lvis evo device will expand and total length may foreshorten up to 60% from its undeployed length (refer to table 1) as it exits the microcatheter.Ensure microcatheter is retracted and clear from the proximal flared ends.Note: visualize and refer to the implant radiopaque end markers to maintain adequate implant length, approximately 7 mm or adequate length on each side of the aneurysm neck or target location to ensure appropriate neck coverage.Warning: do not detach the lvis evo device if it is not properly positioned in the parent vessel.Observe the delivery wire distal tip to assure it remains within the desired location of the parent vessel.21.Prior to removing the delivery wire and if necessary, carefully position the microcatheter distal to the lvis evo device to maintain access through the lvis evo device.Remove and discard the delivery wire.Warning: the lvis evo device delivery wire should not be utilized as a guidewire.Do not torque the lvis evo device.A torque device should not be used.23.Use the guidewire and microcatheter to access the aneurysm through the lvis evo device cells.Warning: observe lvis evo device marker position during placement of the microcatheter into the aneurysm to ensure that the lvis evo device does not migrate or dislodge from its deployed position.Note: access to the aneurysm may be facilitated by the use of a microcatheter that has been shaped.25.Warning: observe lvis evo device marker position during the coiling procedure to ensure that the device does not migrate from its deployed position.After placing the last coil, verify that the lvis evo device has remained patent and properly positioned.Advance a guidewire, if necessary, to the microcatheter tip and carefully remove the microcatheter.Note: a microcatheter may be positioned into the aneurysm sac prior to delivery of the lvis evo device.The microcatheter will be supported by the lvis evo device during delivery of embolic coiling.After completing the coiling, the coiling microcatheter should be carefully removed to avoid dislodging the lvis evo device.27.Caution: carefully watch the lvis evo device distal and proximal markers when passing through the deployed lvis evo device with embolic coiling microcatheters to avoid displacing the lvis evo device.No images of the embolization procedure ((b)(6) 2023) or of the first follow-up (at the time of symptoms, in (b)(6) 2023) were provided.However, two dicom files containing a left vertebral angiogram, a left ica angiogram, and a cerebral mri/mra dated on (b)(6) 2023) were provided for review.The review of these files is as follows: left vertebral: ap and lateral dsa images, subtracted and unsubtracted, with and without contrast.An lvis evo spans from the distal basilar artery to the right p1/p2 junction; it appears well-opened.There may be very mild distal fish mouthing.There are coils in a small basilar tip aneurysm.There is a 59% in-stent stenosis of the entire stent, as shown on an ap image where measurements were taken at the level of the mid-p1: residual lumen: 0.99 mm, stent itself: 2.40mm.There is no flow limitation.Left ica: ap and lateral dsa images, subtracted and unsubtracted, with and without contrast.Normal angiogram.Cerebral mri/mra, (b)(6) 2023): multiple sequences were obtained.The mra redemonstrates the in-stent stenosis; there is no evidence of ischemic infarcts.There is a small susceptibility artifact in the mid right cerebellum, of unknown origin.These images show the presence of the in-stent stenosis; however, they do not explain why the in-stent stenosis occurred.Without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.
|
|
Event Description
|
Updated information in the format of procedural images was received and reviewed.Please see h6 & h10.Other previously submitted information remains unchanged at this time.
|
|