Device Problem Sharp Edges (4013)
Patient Problem Laceration(s) (1946)
Event Date 08/31/2023
Event Type  Injury  
Manufacturer Narrative
Ge healthcares investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Block d4 unique identifier: (b)(4).Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718 h3 other text : device evaluation anticipated, but not yet begun.
 
Event Description
The hospital reported an employee was working to replace the inlay at the bottom of the surgical table, when his right hand came across the sharp edge of the inlay causing a laceration to his right hand.The employee reportedly requested stiches.Ge healthcare will submit a follow-up report when the investigation has been completed.
 
Manufacturer Narrative
Ge healthcareâs (gehc) investigation has been completed.The root cause of the laceration was use error.The gehc employee did not wear cut-resistant gloves during the repair process.Gehc environmental health and safety created a safety communication to create additional awareness that cut-resistant gloves should be worn during the repair process.In addition, as continuous improvement, clarity will be provided in product service manuals for use of protective equipment when performing repairs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS CS2
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key17929719
MDR Text Key325571793
Report Number2112667-2023-05321
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/14/2023
Initial Date FDA Received10/13/2023
Supplement Dates Manufacturer Received03/25/2024
Supplement Dates FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
-
-