PHILIPS MEDICAL SYSTEMS MX40 1.4 GHZ SMART HOPPING Back to Search Results
Model Number 865350
Device Problems Device Alarm System (1012); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2023
Event Type  malfunction  
Event Description
The customer reported that there was a damaged case.In addition, it was stated that the device did not alarm/alert as intended.
 
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Manufacturer Narrative
The philips authorized repair facility evaluated the device and was unable to replicate the reported problem.It was confirmed the speaker produced sound.After testing it was determined that the speaker was functioning as designed.If additional information is received the complaint file will be reopened.
 
Event Description
Philips received a complaint on the mx40 1.4 ghz smart hopping indicating that the mx40 could not provide alarming due to a broken case & unable to hold batteries.It is unknown if the device was in use at time of event, and there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MX40 1.4 GHZ SMART HOPPING
Type of Device
MX40 1.4 GHZ SMART HOPPING
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hauke schik
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key17929961
MDR Text Key325599103
Report Number1218950-2023-00760
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838082236
UDI-Public00884838082236
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number865350
Device Catalogue Number865350
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/15/2023
Initial Date FDA Received10/13/2023
Supplement Dates Manufacturer Received10/17/2023
Supplement Dates FDA Received11/13/2023
Date Device Manufactured07/12/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-