Model Number 865350 |
Device Problems
Device Alarm System (1012); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/15/2023 |
Event Type
malfunction
|
Event Description
|
The customer reported that there was a damaged case.In addition, it was stated that the device did not alarm/alert as intended.
|
|
Manufacturer Narrative
|
A follow up report will be submitted once the investigation is complete.
|
|
Manufacturer Narrative
|
The philips authorized repair facility evaluated the device and was unable to replicate the reported problem.It was confirmed the speaker produced sound.After testing it was determined that the speaker was functioning as designed.If additional information is received the complaint file will be reopened.
|
|
Event Description
|
Philips received a complaint on the mx40 1.4 ghz smart hopping indicating that the mx40 could not provide alarming due to a broken case & unable to hold batteries.It is unknown if the device was in use at time of event, and there was no adverse event reported.
|
|