Catalog Number AB2000
Device Problems Display or Visual Feedback Problem (1184); Visual Prompts will not Clear (2281)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/03/2023
Event Type  malfunction  
Manufacturer Narrative
Component code = 4756 - aquabeam console, a reusable component of the aquabeam robotic system, controls the functionality of the high-velocity waterjet delivered by the aquabeam handpiece.Root cause of reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation (procep) became aware that during the aquablation procedure the aquabeam robotic system generated multiple "e04 -console error" messages, which had to be cleared throughout the procedure to complete three treatment pass.The reported event caused a surgical procedure delay of over 20 minutes.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
The aquabeam console was returned for investigation.Functional testing was able to replicate the e04 error upon priming at 50% and 100% power.E02 and e03 errors were also observed during functional testing.Additional analysis noted oil and debris particulates on the encoder sensor and was identified to be the main cause of the e04 issues.A review of the device history record (dhr) (b)(4) / serial number (b)(6) and aquabeam console / lot number 23c02312 was conducted, which confirmed that there were no non-conformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system met all design and manufacturing specifications when released for distribution.The current user manual um0102-00 rev.D, aquabeam robotic system user manual, canada, english was reviewed.Table 5: system detected errors and faults e04 - console error; release foot pedal and click x.If error persists, turn off and turn on console.The root cause of the reported event has been determined to be due to design.The reported issue is being addressed through our quality system.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
150 baytech drive
san jose CA 95134
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
150 baytech drive
san jose CA 95134
Manufacturer Contact
doria esquivel
150 baytech drive
san jose, CA 95134
6502327291
MDR Report Key17936337
MDR Text Key325699125
Report Number3012977056-2023-00178
Device Sequence Number1
Product Code PZP
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/03/2023
Initial Date FDA Received10/13/2023
Supplement Dates Manufacturer Received01/04/2024
Supplement Dates FDA Received01/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/31/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-