Catalog Number 8065977763
Device Problem Failure to Advance (2524)
Patient Problem Insufficient Information (4580)
Event Date 08/23/2023
Event Type  malfunction  
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history record could not be reviewed because the reporting facility did not provide a valid lot number or any identification traceable to the manufacturing documentation.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event Description
A health care professional reported with a description of iol (intraocular lens) was stuck in the cartridge and was not ejecting.There was no patient harm and the surgery was completed with another lens during the same procedure.
 
Manufacturer Narrative
The product was not returned.The monarch complaint history and product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.Information provided on completed form from the customer indicated that a qualified cartridge and viscoelastic were used.The form indicated "other" for the associated handpiece from the listed company handpieces.This answer would indicate that a non-qualified handpiece was used.The root cause for the reported complaint was most likely related to a failure to follow the ifu.Company foldable iols are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The file information indicated that the sample will not be returned.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17938383
MDR Text Key325687834
Report Number1119421-2023-01753
Device Sequence Number1
Product Code MSS
UDI-Device Identifier00380659777639
UDI-Public00380659777639
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065977763
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/19/2023
Initial Date FDA Received10/16/2023
Supplement Dates Manufacturer Received11/06/2023
Supplement Dates FDA Received11/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ACRYSOF IQ VIVITY EXTENDED VISION IOL; HANDPIECE- OTHER; MONARCH II LOADING FORCEPS; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE
Patient Age50 YR
Patient SexFemale
-
-