The strip lot number and expiration date were requested but were not provided.The meter and strips have been requested for return.The product has not been received at this time.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods." per product labeling: anti-phospholipid antibodies (apa) like lupus antibodies (la) may cause falsely prolong coagulation times.(false-high inr values /false-low %quick values).Where apa are known to be present or suspected, testing with the coaguchek has to be discontinued until discussed with the patient´s physician.E3 - occupation was patient/consumer.
|