MEDTRONIC NAVIGATION, INC O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M
|
Back to Search Results |
|
Model Number 9732719 |
Device Problems
Use of Device Problem (1670); Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/12/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H3, h6: no hardware parts have been returned for analysis.B17, c20, d15 are applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that when the manufacturer representative attempted to turn on the system it sounded like one of the relay switches was bad and it shut itself off after a couple minutes. there was no patient involvement.
|
|
Manufacturer Narrative
|
Concomitant medical products: product id: bi71000581, lot #: (b)(4), product id: bi71000524, lot #: (b)(4), product id: bi31000172r, lot #: (b)(4), product id: bi31000125, lot #: unknown, product id: bi31000126, lot #: unknown, product id: bi71000125, lot #: unknown.H6: the system was serviced in the field.Replaced all x-ray and motion batteries, replaced all 3 charger boards and the pfc1000 board.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them, because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank, because the information is currently, unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3,h6: a hardware analysis was initiated for bi71000524 to determine the probable cause of the reported behavior.Analysis found that there were no failures found and the complaint was not verified.The battery charger pcbas passed the bench test and visual inspection.Chargers output voltages were within inspection parameters.Code b01 is applicable and previously reported.Codes c19,d14 are applicable.H3,h6: a hardware analysis was initiated for bi31000172r to determine the probable cause of the reported behavior.Analysis found that there were no failures found and the complaint was not verified.The pfc board passed visual inspection and the bench test.110vac was used to power throughout the board while the boards output was 380vac, which was within the correct output spec.When 24vac was applied to the fan, the fan was fully functional.Code b01 is applicable and previously reported.Codes c19,d14 are applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Continuation of d11: section d information references the main component of the system.Other relevant device(s) are: product id: bi71000581, lot/serial #: (b)(6) rev.B h3) the motion battery charger was returned to the manufacture for analysis.The battery charger passed bench test and visual inspection.Chargers output voltages were within inspection parameters.Installed battery charger in a known functional system.The system initialized.Motion, generator, communication and charging readied.2d and 3d images readied.No failure found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|