Model Number M00546610
Device Problems Difficult to Advance (2920); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2023
Event Type  malfunction  
Manufacturer Narrative
Block h6: imdrf device code a180104 captures the reportable event of presence of foreign material at the tip of the spyscope.
 
Event Description
It was reported to boston scientific corporation that a spyscope ds ii access & delivery catheter was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed for the treatment of bile duct stricture on (b)(6) 2023.During the procedure, the physician decided to used spyglass after performing intraductal radiofrequency ablation to see what the tissue looked like post-ablation.The physician felt resistance during advancing the spyscope ds ii down the working channel of the duodenoscope, and they saw on the spy screen what appeared to be a metal filament at the tip of the spyscope.They withdrew both scope from the patient, and able to identify the metal filament; however, it was unclear where the metal was coming from.Reportedly, there were no problems passing any device down the duodenoscope prior using the spyscope ds ii.The procedure was completed with another of the same device.There were no patient complications reported as a result after this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPYSCOPE DS II ACCESS & DELIVERY CATHETER
Type of Device
CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17941601
MDR Text Key326394611
Report Number3005099803-2023-05540
Device Sequence Number1
Product Code FBN
UDI-Device Identifier08714729965404
UDI-Public08714729965404
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00546610
Device Catalogue Number4661
Device Lot Number0031497945
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/21/2023
Initial Date FDA Received10/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-