The product investigation was completed.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.Visual inspection and functional tests of the returned device were performed following bwi procedures.Visual analysis uncovered a hole and reddish material within the pebax; however, no other abnormalities were identified.Subsequent temperature and impedance tests were conducted, but no temperature readings were displayed due to an open circuit in the tip area.The damage on pebax and reddish material observed could be related to the open circuit founded, and its associated with the event reported by the customer.The root cause may be linked to the handling process, as there are control inspection points in place to prevent such issues.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The instructions for use contain the following warning stated in the carto 3 system manual: if the radiofrequency (rf) generator does not display temperature, verify that the appropriate cable is plugged into the rf generator.A manufacturing record evaluation was performed for the finished device 31047183la number, and no internal actions related to the reported complaint condition were identified.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
It was reported that a patient underwent an ischemic ventricular tachycardia (isvt) ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and post procedure the bwi product analysis lab identified a hole in the pebax.During the procedure, an error occurred on the smartablate generator stating that "ablation temp exceeds limit" and ablation was not possible when the physician attempted rf (radiofrequency) delivery.The catheter was removed from the patient and inspected, and then flushed without issue.It was reinserted and ablation attempted again with the same result.The generator was rebooted and cable was exchanged without resolution.The catheter was exchanged and the issue resolved, case continued successfully.The temperature issue is not mdr-reportable.The hole in the pebax is mdr-reportable.
|