• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 IV; FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 IV; FREE THYROXINE Back to Search Results
Catalog Number 09043284190
Device Problem Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/25/2023
Event Type  malfunction  
Manufacturer Narrative
The cobas 8000 core unit serial number was (b)(6).The investigation is ongoing.
 
Event Description
There was an allegation of questionable ft4 g4 elecsys results from the cobas 8000 core unit.The sample was retested on wako accuraseed assay and abbott alinity assay.The sample was submitted for preliminary investigation and was tested on a cobas e 801 module.Refer to the attachment to the medwatch for all patient data.No information was provided to determine if the questionable result was reported outside of the laboratory.
 
Manufacturer Narrative
The investigation could not identify a product problem.Assays from different vendors can generate different values.This relates to the overall setups of the assays, the antibodies used, differences in reference materials/methods, and differences in the standardization methodology used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 IV
Type of Device
FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17942501
MDR Text Key325760414
Report Number1823260-2023-03313
Device Sequence Number1
Product Code CEC
UDI-Device Identifier07613336172780
UDI-Public07613336172780
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
K220456
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number09043284190
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/02/2023
Initial Date FDA Received10/16/2023
Supplement Dates Manufacturer Received10/27/2023
Supplement Dates FDA Received10/27/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age50 YR
Patient SexMale
-
-